User`s manual
Operation Theory   53
ADLINK Technology Inc. PCI-9524 
Copyright 2008 User’s Manual
Table 4-1: Data Rates vs. Multiplexing, Auto-zero & 
ADC Sampling Rates, in Samples-per-second (SPS)
Non-multiplexed  Multiplexed ADC 
Sampling 
Rate 
Auto-zero 
Disabled 
Auto-zero 
Enabled 
Auto-zero 
Disabled 
Auto-zero 
Enabled 
30,000 
30,000 
818.73 1637.47  818.73 
15,000 
15,000 
768.40 1536.81  768.40 
7,500 
7,500 
703.53 1407.06  703.53 
3,750 
3,750 
594.74 1189.48  594.74 
2,000 
2,000 
462.66 925.33  462.66 
1,000 
1,000 
316.32 632.63  316.32 
500 
500 
193.75 387.49  193.75 
100 
100 
47.26** 94.51  47.26** 
60* 
60 
29.00 58.00  29.00 
50** 
50 
24.29 48.59  24.29 
30* 
30 
14.74 29.49  14.74 
25** 
25 
12.32 24.64  12.32 
15* 
15 
7.44 14.87  7.44 
10*** 
10 
4.97 9.94  4.97 
5*** 
5 
2.49 4.99  2.49 
2.5*** 
2.5 
1.25 2.5  1.25 
4.2.5 Power Line Noise Rejection
The SINC filter built into the PCI-9524 works best for suppressing 
power line noise, if the ADC sampling rate is set to match power 
line frequency. The harmonics of the power line noise can also be 
suppressed as well, see Table 4-3 for illustration. For applications 
demanding high-stability, low-drifting measurements, selecting a 
sampling rate that provides inherent power line noise rejection is 
recommended.
NOTE:
NOTE:
1) For the equivalent data rate per channel, divide the multi-
plexed data rate figures by four.
2) *60 Hz Rejection, **50 Hz Rejection, ***Simultaneous 50 
and 60 Hz Rejection.
3) Auto-zero function is always disabled for general purpose 
input channels, i.e. channels 4 to 7.