Manual
28
•
Calibration
5.2 VR Assignment
There are twelve variable resistors (VR) on the ACL-6128A board to allow
users to make accurate adjustment on two D/A channels. VR1 to VR6 belong
to CH1, and VR7 to VR12 belongs to CH2. The detailed functionality of each
VR is listed below:
CH1‘s VR
VR1
(-10V) Internal reference
VR2
(-5V) Internal reference
VR3
Gain Adjustment
VR4
Unipolar Offset
VR5
Current Sink Offset (4mA)
VR6
Bipolar Offset
CH2‘s VR
VR7
(-10V) Internal reference
VR8
(-5V) Internal reference
VR9
Gain Adjustment
VR10
Unipolar Offset
VR11
Current Sink Offset (4mA)
VR12
Bipolar Offset
There are two testing points (TP1 and TP2) on the board, which are used for
calibration the ACL-6128A. TP1 is for CH1, and TP2 is for CH2.
5.3 Internal Reference Source Adjustment
The Internal Reference Source adjustment can ensure that the internal
reference voltages of the ACL-6128A can offer very accurate voltage source—-
5V and -10V.
1.
Ground the black probe of the voltmeter.
2.
Connect the other probe to TP1 (for testing CH1), or TP2 (for testing
CH2).
Change the jumper to JP3 (for CH1) or JP7 (for CH2), trim
VR1
(for
CH1) until the voltmeter reads
-10V
, trim
VR7
(for CH2) until the
voltmeter also reads
-10V
.
3.
Change the jumper to JP3 (for CH1) or JP7 (for CH2), trim
VR2
(for
CH1) until the voltmeter reads
-5V
, trim
VR8
(for CH2) until the
voltmeter also reads
-5V
.