Manual
98 • Calibration
Table 5-2: ND-6013 Calibration Resistance
Code Input Range
Span Calibration
Resistance
Offset Calibration
Resistance
20
Pt-100, -100°C to
+100°C, α=.00385
200Ω 50Ω
21
Pt-100, 0°C to +100°C,
α=.00385
200Ω 50Ω
22
Pt-100, 0°C to +200°C,
α=.00385
200Ω 50Ω
23
Pt-100, 0°C to +600°C,
α=.00385
350Ω 50Ω
24
Pt-100, -100°C to
+100°C, α=.003916
200Ω 50Ω
25
Pt-100, 0°C to +100°C,
α=.003916
200Ω 50Ω
26
Pt-100, 0°C to +200°C,
α=.003916
200Ω 50Ω
27
Pt-100, 0°C to +600°C,
α=.003916
350Ω 50Ω
28
Ni-100, 0°C to +100°C 200Ω 50Ω
29
Ni-120, 0°C to +100°C 350Ω 50Ω
2A 0 Ohm to 60Ohm
60Ω 0Ω