Command Set 49
3.12 Span Calibration to each Channel
@Description
To correct the gain errors of AD converter by using the span calibration.
@Syntax
$(Addr)0(Channel No)<CR>
$ Command leading code (1 character)
(Addr) Address ID (2 character)
0 Span calibration (1 character)
(Channel No) Channel for Calibration (1 character) 0~2
@Response
!(Addr)<CR>
or
?(Addr)<CR
>
!
Command is valid.
?
Command is invalid.
(Addr)
Address ID.
@Example
User command: $0601<CR>
Response: !06<CR>
To perform the span calibration for analog input module ND-6013
channel 1, address ID is 06H.
Note: To perform the calibration, a proper input signal should be
connected to the analog input module. Different input range
have different input voltage, detail refer chapter 5
Calibration” .
(for 6013 F/W version
A
3.05~A4.60)