Owner manual
Calibration • 97
Analog Input Module‘s Calibration Voltages
Table 5-1: ND-6011/D/ND-6018 Calibration voltages
Code Input Range
Offset Calibration
voltage
Span Calibration
Voltage
00
±15 mV
0 mV +15 mV
01
±50 mV
0 mV +50 mV
02
±100 mV
0 mV +100 mV
03
±500 mV
0 mV +500 mV
04
±1 V
0 V +1 V
05
±2.5 V
0 V +2.5 V
06
±20 mA
0 mA +20 mA
0E
J (0°C to 760°C)
0 mV +50 mV
0F
K (0°C to 1000°C)
0 mV +55 mV
10
T (-100°C to 400°C)
0 mV +30 mV
11
E (0°C to 1000°C)
0 mV +78 mV
12
R (500°C to 1750°C)
0 mV +25 mV
13
S (500°C to 1750°C)
0 mV +19.5 mV
14
B (500°C to 1800°C)
0 mV +15 mV
15
N (-270°C to 1300°C)
0 mV +55 mV
16
C (0°C to 2320°C)
0 mV +39 mV
Table 5-1: ND-6012/D/ND-6017 Calibration voltages
Code Input Range
Offset Calibration
voltage
Span Calibration
Voltage
08
±10 V
0 mV +10 V
09
±5 V
0 mV +5 V
0A
±1 V
0 mV +1 V
0B
±500 mV
0 mV +500 mV
0C
±150 mV
0 mV +150 mV
0D
±20 mA
0 mA +20 mA