5-2 ADS TRITON+ Manual
Installation
A typical field installation involving external power requires the
following:
An AC mains power line terminated at a pole or post close to
the manhole in which the monitor is installed
An AC distribution device (i.e., disconnect, fuse, or breaker
box) on the pole/post
An earth ground rod installed next to the pole/post bonded to
the AC distribution device
An AC to 24-volt DC converter (i.e., power supply). ADS
recommends a power supply available through ADS (p/n
508293), which works with worldwide AC power (94 240
VAC, 50 60 Hz). This power supply is provided
automatically as part of the XIO and XBUS, but must be
ordered for the ExPAC, when a power supply is required.
However, please note that the customer may use another AC-to-
DC converter/power supply, if desired, to support the ExPAC,
provided it can supply from 9 to 36 volts of power at a
minimum of 15 watts.
A power cable running between the external DC source (i.e.,
power supply) and the ExPAC power input. The customer
must supply this cabling of the appropriate length and
specifications
A properly rated enclosure to house and protect the ExPAC and
power supply
Note: Applications involving an XIO or XBUS do not
require an additional power cable or an enclosure because
they are included. The power is pre-run between the
power supply and ExPAC inside the XIO enclosure.
A communication cable (ADS p/n 8000-0378-xx, where xx
represents the required length of the cable) running between
the ExPAC/XBUS/XIO and the monitor