Configura tion and Activation 6-75
2. Allow the TRITON+ monitor to log data for an appropriate
amount of time to obtain an ad eq uate amount of d ata. Please
note the interval at which the monitor takes readings for
Modbus output is based on the Scan rate.
3. Collect the data fro m the monitor. Refer to Collectin g Data
from the Monitor on page 6-82 for detailed instructio ns.
4. Compare the data colle c te d from the monitor to the output
values obtained thr ough Modbus. The data collected through
Qstart reflects an average of the readings taken over a specific
interval; the values retrieved through the Mo dbus interface
represent the cur r ent (persistent) data in the monitor from the
last interval (readings). Therefore, the values compared may
not be exactly the same, but should be similar.