Running Sensor Diagnostics
Qstart’s diagnostics tool enables the user to verify the prope r
operation of sensors, obtain current readings and status, adjust
settings, and identify, diagnose, and troubleshoot potential problems
with ultrasonic depth, velocity, and p r essure sensors. Run
diagnostics for the system sensors in the following way:
Note: Device diagnostics for TRITON+ monitors
equipped with SIMs with dynamic IP addresses is only
available when connected serially.
Perform diagnostics on a de vice in the following wa y:
1. Select the monitor for which you want to run diagnostic s on a
sensor from the Location Name field in the Monitor section
on the ADS Qstart dialog.
2. Click on the Connect button in the Functions section.
Qstart initiates and establishes communication with the
monitor. The status bar displays Ready once a connection has
been successfully made.
3. Select the Monitoring Point 1 or Monitoring Point 2 tab
which contains the device you want to diagnose.
4. Select the device rep r esenting the individual sensor(s) on which
you want to perform diagnostics from the Devices list.
5. Click on the View button.
The device dialog corresponding to the selected device
displays.
6. When a pplicable, click on the tab corresponding to the
particular sensor on which you want to run diagnostics.
7. Click on the Read button be low the Results window.