Problem
Modem connec ts but monitor does not respond.
Possible
Cause
Connection t ype configurat i on has reverted to Serial.
3.3-volt fuse on regulator may have blown.
Possible
Solutions
Verify t hat the com municat i on type is set to Wireless.
Connect direc tly with the moni t or using a USB Serial
Interface cable (ADS p/n 8000-0337). Verify the monitor
responds. If it does , the modem might be faulty. If it
does not, remove the m oni tor from t he hazardous area
(at least 20 feet away from the manhole), disconnect
the battery pac k, wait at least 1 minut e, and then
reconnect the monit or bat tery pack. If remote
com municat i on res umes, t ry t o connect several times t o
ensure communication is reli abl e.
Replace the 3.3-volt yellow fuse (when replac eabl e) on
the regulator.
Contact your regional ADS representati ve.