The following tables contain gene ral tr oubles hooting t echniques for
the velocity subsystem corresponding to the Peak Combo Sensor
and Surface Combo Senso r ( Surcharge Velocity Sensor only).
Since the Surface Combo Sensor is mounted at the top of the pipe,
potential issues involving a sensor that is out of the flow d o not
apply to the Surface Combo Sensor.
Note: Contact your regi onal ADS representative for
further diagnosis prior to replacing a sensor.
Problem
Monitor often provides a vel ocity reading of 0.
Possible
Causes
Sensor’s velocity t ransmit / receive surfaces may be dirt y
or the veloci ty function i s broken.
Minimum flows m ay be oc curring within the sens or
deadband (extending 1.0 inc hes (25 mm) up from the
bottom of the Peak Combo Sensor).
Cabling between the proces sor board and the veloci t y
sensor may be bad.
Veloci t y parameters may require adjustment.
Possible
Solutions
Clean the sensor’s trans mit/ receive surfaces.
Verify that minimum flows are occurring outside the
sensor deadband.
Check the vel oc i ty parameters .
Inspec t the sensor c abl es for tightness.
Check for moist ure i n t he connector.
Contact your regional ADS representative.
Replace the sensor.
Replace the monitor if defective.