User Manual

8-28 ADS FlowShark IS Manual
Problem
Monitor often provides a vel ocity sensor reading of 0.
Possible
Causes
Sensor m ay be di rt y or brok en.
Cabling between the analog board and the sensor may
be bad.
Velocity parameters may require adjus tment.
Possible
Solutions
Clean the sens or.
Check the vel oc i ty paramet ers.
Inspec t the sensor cables f or t i ghtness.
Check for mois ture in the c onnector.
Contact your regional ADS represent at i ve.
Replace the s ensor.
Replace the monitor if defective.
Problem
Velocity sensor data does not fluc t uate much (but is not
0).
Possible
Cause
Sensor m ay be di rt y
Sensor may be broken.
Possible
Solutions
Clean the sens or.
Check the sensor connections .
Check the vel ocity parameters.
Contact your regional ADS represent at i ve.
Replace the sensor.