Note: With the battery packs connected and the dip switches ON,
the PrimeProbe2 is now functional. The next step is to configure
the converter corresponding to the specific application.
Connecting Mains Power
Important Precautions to Consider Before Connecting to Mains
Power
Only qualified technicians with appropriate knowledge of local
codes and practices should make connections to mains power.
Verify the mains voltage falls within the limits stated on the tag
plate.
Use appropriate fireproof wire and connectors in accordance with
local code.
Equip mains power supply line with external protection for current
overload (i.e., fuse or automatic line breaker with limiting capacity
of less than or equal to 3A @ 240V).
Circuit breaker should be located within close proximity to the
PrimeProbe converter.
1. Shut off the mains power supply at the closest circuit breaker.
2. Remove the back from the PrimeProbe Converter.
3. Loosen the topmost gland on the side of the converter.
4. Slide the mains cable through the gland and into the
termination area of the converter.