User Manual

PrimeProbe2 User Manual
4
Technical Specifications
Parameter Specification
Pipe Sizes
3.2 to 315.0 inches (80 mm to 8000 mm) at 1/8 pipe
diameter (insertion depth) installation
Measurement Range
Bi-directional from 0.03 to 16.40 feet per second (0.01 m/s
to 5 m/s) (Maximum may be lower based on insertion length
and pipe position.)
Accuracy; Point
Velocity
Flow
±0.0066 feet per second (± 2mm/sec) or ± 2%, whichever is
the greater.
Refer to ISO 7145-1982
Repeatability
± 0.5% of velocity
Flow Determination
Assumes fully developed flow profile; ISO 7145-1982
reference
Measurement
Sampling
Continuous or programmable from 1 to 90 seconds
Minimum Fluid
Conductivity
5µS/cm
Process Connection
1-inch (25-mm) BSP (British Standard Pipe)
Pipeline Pressure
Rating
232 psi
Sensor Material
PTFE (polytetrafluoroethylene)
Body Material
SS304
Electrode Materials
SS316
Flow Temperature
32º to 176º F (0 to 80º C)
Safety
Probe fitted with anti-bounce chain
Pressure Tapping
Female quick-release connector
Adjustment Method
5 mm Allen key (supplied by ADS) fits all screws
Protection
IP68
Batteries
Type: Lithium batteries
Life: 3 years (depending on sampling period and usage)
Output: Signal
Units
Connection
Pulses proportional to velocity/flow rate - maximum frequency
32 Hz
User-selectable
MIL-spec connector
Communications
USB