User's Manual

18
5. Remove the token from the token slot on the auxiliary display.
6. By repeating steps 4 and 5, as many as two additional auxiliary displays
may be linked. All linking must be done within two minutes.
7. Store the token in the transmitter token slot when not in use.
Note: If multiple transmitters are within 100 feet of each other, for example, inadjacent
operating rooms, link each transmitter to its respective display(s) one set at a time. Wait 15
seconds before linking the next transmitter/display set.
Permanent Linking
Please contact your Stryker rep in order to permanently link the display(s)
with a specic transmitter, as deemed appropriate. A permanent link may
be desirable in certain cases. Once the display(s)/transmitter set has been
permanently linked, the set will automatically link when powered on.
Display(s)/transmitter sets that have been set to permanent link mode should
be kept in the same operating room.
To return the display(s)/transmitter set to default linking mode please follow
the instructions below using a blue WiSe Linking Token.
1. Power on the display(s) and transmitter.
2. Insert the blue token into the transmitter and remove it when the token
LED turns green.
3. Insert the blue token into each display, one at a time, and remove the token
when the token LED turns green.
4. Power o all devices.
5. Power on all devices.
Now the system can be linked as described in the default linking mode
procedure.