User guide

36 User Guide for the Neptune™ Sound Processor
Recommended Neptune System Operating and Storage Temperatures
Table 4: Operating and storage temperature ranges
Condition Minimum Maximum
Operating temperature 0°C (32°F) 45°C (115°F)
Storage temperature -20°C (-4°F) 55°C (131°F)
The Neptune system meets the temperature safety requirements of IEC 60601-
1:2005 in ambient temperatures up to 36.8°C (98°F).
Approved Power Sources
Advanced Bionics has verified Neptune function with the power sources listed
below; behavior when using power sources other than those listed here cannot be
guaranteed.
Table 5: Approved Neptune power sources by geographic region
Power Source Manufacturer Model Number
NORTH AMERICA
AAA Battery Energizer ® E92 (Alkaline)
Energizer L92 (Lithium)
Duracell ® DC2400 (NiMH)
Programming Interface
(used during fitting, only) Advanced Bionics AB-6500
EU
AAA Battery Energizer Ultra Plus LR03 (Alkaline)
Energizer L92-FR03 (Lithium)
Duracell HR03 (NiMH)
Programming Interface
(used during fitting, only) Advanced Bionics AB-6500
Table 6: Approved Neptune power supplies
Power Supply Rated Voltage Type Rated Current
AAA Battery 1.0-5.0 V DC 250mA
Programming Interface 3.1-5.0 V DC 250mA