1.3 Cleaning/Disinfecting
During normal use of the POC-175 may become soiled and should, there-
fore, be cleaned regularly. Agents: Green tinctured soap and Enzymatic
detergents
Steps:
1. Wipe the POC-175 with a clean cloth that has been moistened in
the cleaning solution.
2. Prepare agent per manufacturer’s instructions or hospital protocol.
3. Wipe thoroughly with a clean cloth
1.4 LCD Specifications
Display type: 17" TFT LCD
Resolution: 1280 x 1024
Colors: 16.2M (6 bits + FRC)
Dot size (mm): 0.264 x 0.264
Viewing angle: 140/130°
Luminance: 300 cd/m2
Contrast ratio: 500:1
LCD MTBF: 50,000 hours
Backlight lifetime: 50,000 hours
Cautions Do not immerse or rinse the POC-175 and its periph-
erals. If you accidentally spill liquid on the device, dis-
connect the unit from the power source. Contact your
Biomed regarding the continued safety of the unit
before placing it back in operation.
Do not spray cleaning agent on the chassis.
Do not use disinfectants that contain phenol.
Do not autoclave or clean the POC-175 or its periph-
erals with strong aromatic, chlorinated, ketone, ether,
or Esther solvents, sharp tools or abrasives. Never
immerse electrical connectors in water or other liq-
uids.