User manual
POC-S195 User Manual 4
1.2.9 Environment
Temperature: 0 ~ 35° C (32 ~ 95° F)
Relative humidity: -5° C to 40° C / 95% RH operating, -40° C to 60° C/ 95%
RH Storage
Shock: 20 G, half sine, 11 msec duration
Vibration: 5-500Hz, PSD 0.002G²/Hz, 1Grms, 1 Oct/min, 3 axes, 1 hr/per axis
Altitudes: Operational: 6,000 feet; shipping: 40,000 feet
Certifications: EMC: CE, FCC approved
Safety: UL60601-1 and EN60601-1 approved.
This device bears the CE label in accordance with the provisions of the EMC
Directive 89/336/EMC and the Low Voltage Directive 73/23/EEC.
1.2.10 Cleaning/Disinfecting
During normal use, the POC-S195 may become soiled and should, therefore, be
cleaned regularly. Agents: Green tinctured soap and Enzymatic detergents
Steps:
1. Wipe the POC-S195 with a clean cloth that has been moistened in the cleaning
solution.
2. Prepare agent per manufacturer's instructions or hospital protocol.
3. Wipe thoroughly with a clean cloth
1.3 1.3 LCD Specifications
Display type: 19" TFT LCD.
Resolution: 1280 x 1024
Colors: 16.7M (8 bits/color)
Dot size (mm): 0.294 x 0.294
Viewing angle: 178°x178°
Luminance: 300 cd/m2
Contrast ratio: 1300 : 1
LCD MTBF: 50,000 hours
Backlight lifetime: 50,000 hours
Caution! Do not immerse or rinse the POC-S195 and its peripherals. If you acci-
dentally spill liquid on the device, disconnect the unit from the power
source. Contact your Biomed regarding the continued safety of the unit
before placing it back in operation.
Do not spray cleaning agents on the chassis.
Do not use disinfectants that contain phenol. Do not autoclave or clean
the POC-S195 or its peripherals with strong aromatic, chlorinated,
ketone, ether, or Esther solvents, sharp tools or abrasives. Never
immerse electrical connectors in water or other liquids.