Instruction Manual

68
Technical
Data
Temperature-Measurement range:
Measurements in the “ear area” : 34~42.2°C (93.2~108°F),
Scan- operation: -22~80°C (-7.6~176°F).
Operation temperature range: 10~40°C (50~104°F), relative
humidity ≤85%
Storage temperature range: -20~50°C (-4~122°F), relative
humidity ≤85%.
Transport temperature should be lower than 70°C, relative
humidity ≤95%.
Atmospheric pressure: 800~1013 hPa
In accordance to ASTM E1965-98, EN12470-5:2003
Clinical Thermometers Part 5: Performance of Infra-red ear
thermometers (with maximum device), IEC/EN60601-1-2
(EMC), IEC/EN60601-1 Safety Guidlines.
Deviations
With measurements in the ear: +/-0.2°C (0.4°F) within the
range of 35.5~42°C (95.9~107.6°F),
+/-0.3°C (0.5°F) out of this range.
In Scan-mode +/-0.3°C (0.5°F) within the range of
22~42.2°C (71.6~108°F),
+/-4% or +/-2°C (4°F) out of this range—
the higher value is decisive.
Fever Alarm saving locations for ear operation (display of
date/time/operation mode for each measurement).
Time display, °C / °F toggle function.
Battery, 1 Lithium cell (CR2032 1 pc.).
Battery life: around 3,000 readings. Enable normal use: 1 year
Expected Service Life: 4 years
Measurement in the ear, the device will calculate the
corresponding value which would be derived from an oral
measurement.
Battery low warning.
Automatic switch-off of measurement activation.
Enclosure Rating: IP22
Dimensions: 146.5 x 38.2 x 54.7 mm
Weight: 55.5 grams including battery
The device should not submerge into any liquids and
expose it to direct moisture.
This symbol indicates important for persons or the
device!
FT 4919 AEG Ear Thermometer (TH520B)
English