Taking Temperature in Ear Temperature Mode
1) When the system is in the state of time/environmental tem-
perature display, press the button of scan, and the last tempera-
ture measurement will be displayed; then after “Beep-Beep“,
the system enters into the mode of measurement.
2)Removetheprobecap,andthethermometerentersinto
theearmeasuringmode(RefertothePicture7.1).
3) Carefully pull back the ear to stretch the auditory canal.
When measuring ear temperature, insert the probe into
the ear rst (refer to the Picture 7.2), then press the but-
ton of scan. One second later you will hear “Beep“, and
you can remove the thermometer. The measurement is
completed.
4)Whenthemeasuredtemperatureexceeds38.0°Cthethermometerwill
send out “Beep-Beep-Beep“ a long one and two short ones, and the
displaywillshowtheiconoffever(RefertothePicture7.3).
Notes:
• Keeptheprobesurfaceclean,otherwisethemeasurementcan‘tbe
accurate.
• Cleantheprobewithcleanclothorpapermoistenedwithwateror
alcohol, and conduct measurement after the water on the surface of
the probe all evaporates.
(Picture 7.1)
(Picture 7.2)
(Picture 7.3)