Installation manual

Model 8310-XX-X-XN IM-493
Aeroflex / Weinschel
21
Change Setup:
0 Server
1 Channel 1
3 E-mail
5 Expert
6 Security
7 Defaults
8 Exit without save
9 Save and exit Your choice ?
Parameters stored ...
Turn off power. To verify the settings you can use the Xport Monitor Mode. To enter Monitor mode, hold down the 'z'
key while powering up the unit. After a few seconds, you should see
*** NodeSet 2.0 ***
0>zzzzzzzzzzzz
9>
Enter 'NC' in uppercase letters, which is the Network Connection command, whose function is to display the unit's IP
configuration
9>NC
IP 010.100.103.049 GW 010.100.103.254 Mask 255.255.255.000 NS 010.100.103.173
0>
NOTE
The parameters shown above were obtained with the Xport set for DHCP mode (IP=0.0.0.0)
Enter 'QU' to quit and exit diagnostics monitor mode.
0> QU
Place the MODE SELECT switch to the ETHERNET position.
To verify network connectivity, you can open a command prompt window on the PC (Start menu | Run… | CMD) and
ping the assigned IP address.
C:\WINDOWS>ping 10.100.103.49
Pinging 10.100.103.49 with 32 bytes of data:
Reply from 10.100.103.49: bytes=32 time=7ms TTL=64
Reply from 10.100.103.49: bytes=32 time<1ms TTL=64
Reply from 10.100.103.49: bytes=32 time<1ms TTL=64
Reply from 10.100.103.49: bytes=32 time<1ms TTL=64
Ping statistics for 10.100.103.49:
Packets: Sent = 4, Received = 4, Lost = 0 (0% loss),
Approximate round trip times in milli-seconds:
Minimum = 0ms, Maximum = 7ms, Average = 1ms