Recommended separation distances between portable and mobile RF 
communication equipment and the Aeroneb Go
The Aeroneb Go nebulizer system is intended for use in the electromagnetic 
environment in which radiated RF disturbances are controlled. The customer 
or the user of the Aeroneb Go can help prevent electromagnetic interference 
by maintaining a minimum distance between port and mobile RF 
communications equipment (transmitters) and the Aeroneb Go nebulizer 
system as recommended below, according to the maximum output power 
of the communications equipment.
Rated maximum 
output power of 
transmitter
W
Separation distance according to frequency of 
transmitter
m
150 kHz to 
80 MHz
d = [1.17] √P
80 MHz to 
800 MHz
d = [1.17] √P
800 MHz to 
2.5 GHz
d = [2.33] √P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.75
1 1.17 1.17 2.33
10 3.70 3.70 7.36
100 11.70 11.70 23.30
For transmitters rated at a maximum output power not listed above, the 
recommended separation distance d in metres (m) can be estimated using the 
equation applicable to the frequency of the transmitter, where P is the maximum 
output power rating of the transmitter in watts (w) according to the transmitter 
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher 
frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic 
propagation is affected by absorption and reection from structures, objects 
and people.