Guidance and manufacturer’s declaration – electromagnetic immunity
The Aeroneb Go nebulizer system is intended for use in the electromagnetic 
environment specied below. The customer or the user of the Aeroneb Go 
should assure that it is used in such an environment.
Immunity test IEC/EN 60601 
Test level
Compliance 
level
Electromagnetic 
environment – 
guidance
Electrostatic 
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be 
wood, concrete or 
ceramic tile. If oors 
are covered with 
synthetic material, 
the relative humidity 
should be at 
least 30%.
Electrical fast 
Transient/burst
IEC 61000-4-4
±2 kV for power 
supply lines
±1 kV for input/
output lines
±2 kV for power 
supply lines
±1 kV for input/
output lines
Mains power quality 
should be that of a 
typical commercial 
or hospital 
environment.
Surge
IEC 61000-4-5
±1 kV differential 
mode
±2 kV common 
mode
±1 kV differential 
mode
±2 kV common 
mode
Mains power quality 
should be that of a 
typical commercial 
or hospital 
environment.
Voltage 
dips, short 
interruptions and 
voltage variations 
on power supply 
input lines
IEC 61000-4-11
<5% Ut 
(>95% dip in Ut) 
for 0.5 cycle
40% Ut 
(60% dip in Ut)  
for 5 cycles
70% Ut 
(30% dip in Ut)  
for 25 cycles
<5% Ut 
(>95% dip in Ut) 
for 5 sec
<5% Ut 
(>95% dip in Ut) 
for 0.5 cycle
40% Ut 
(60% dip in Ut)  
for 5 cycles
70% Ut 
(30% dip in Ut)  
for 25 cycles
<5% Ut  
(>95% dip in Ut) 
for 5 sec
Mains power 
quality should be 
that of a typical 
commercial 
or hospital 
environment. 
If the user of 
the Aeroneb Go 
requires continued 
operation during 
power mains 
interruption, it is 
recommended that 
the Aeroneb Go 
be powered from 
an uninterruptible 
power supply or 
battery.