Immunity test IEC/EN 60601 
Test level
Compliance level Electromagnetic 
environment – 
guidance
Power 
frequency 
(50/60 Hz)
Magnetic eld
IEC 61000-4-8
3 A/m 3 A/m
Power frequency 
magnetic elds 
should be at levels 
characteristic of 
a typical location 
in a typical 
commercial 
or hospital 
environment.
Note: Ut is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The Aeroneb Go nebulizer system is intended for use in the electromagnetic 
environment specied below. The customer or the user of the Aeroneb Go 
should assure that it is used in such an environment.
Immunity 
test
IEC/EN 60601 
Test level
Compliance 
level
Electromagnetic environment – 
guidance
Conducted 
RF 
IEC 
61000-4-6
Radiated RF 
IEC 
61000-4-3
3 Vrms  
150 kHz to 
80 MHz
3 Vrms  
80 MHz to 
2.5 GHz
[3]V
[3]V
Portable and mobile RF 
communications equipment 
should be used no closer to any 
part of the Aeroneb Go, including 
cables, than the recommended 
separation distance calculated 
from the equation applicable to 
the frequency of the transmitter. 
Recommended separation 
distance
d = [1.17] √P
d = [1.17] √P... 80 MHz to 800 MHz
d = [2.33] √P... 800 MHz to 2.5 GHz
where P is the maximum output 
power rating of the transmitter 
in Watts (W) according to the 
transmitter manufacturer and d 
is the recommended separation 
distance in metres (m).