Instruction manual
Guidance and manufacturer’s declaration – electromagnetic immunity
The Aeroneb Go nebulizer system is intended for use in the electromagnetic 
environment specied below. The customer or the user of the Aeroneb Go 
should assure that it is used in such an environment.
Immunity 
test
IEC/EN 60601 
Test level
Compliance 
level
Electromagnetic environment – 
guidance
Field strengths from xed RF 
transmitters, as determined by 
an electromagnetic site survey, ª 
should be less than the compliance 
level in each frequency range. 
b
Interference may occur in the 
vicinity of equipment marked with 
the following symbol:
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher 
frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic 
propagation is affected by absorption and reection from structures, objects 
and people.
ª  Field strengths from xed transmitters, such as base stations for radio 
(cellular/cordless) telephones and land mobile radios, amateur radio, AM 
and FM radio broadcast and TV broadcast cannot be predicted theoretically 
with accuracy. To assess the electromagnetic environment due to xed RF 
transmitters, an electromagnetic site survey should be considered. If the 
measured eld strength in the location in which the Aeroneb Go is used 
exceeds the applicable RF compliance level above, the Aeroneb Go should 
be observed to verify normal operation. If abnormal performance is observed, 
additional measures may be necessary, such as re-orientating or relocating 
the Aeroneb Go nebulizer system.
b
  Over the frequency range 150 kHz to 80 MHz, eld strengths should be less 
than [3]V/m