Instruction manual

TO OPERATE WITH A BATTERY POWERED CONTROLLER
1. Turn the Battery Powered Controller 
upside down. Slide the battery cover 
in the direction of the arrow to 
remove the lid. Put in three (3) "AA" 
disposable alkaline or lithium batteries,
matching them to the plus (+) and 
minus (-) markings inside the Battery 
Powered Controller. Replace the cover 
on the Battery Powered Controller and 
slide the cover until it locks in place.
You can expect at least three (3) hours 
of operation from a new set of 
disposable batteries, (up to five (5) 
hours from high performance batteries).
2. Insert the nebulizer cable connector 
into the back of the Nebulizer Unit.
Slide the cable straight up the 
nebulizer body.
3. Connect the nebulizer cable to the 
Aeroneb
®
Go Battery Powered 
Controller.
4. Turn on the unit by pressing the "on/off" 
button on the Battery Powered 
Controller. The green light on the Battery
Powered Controller will light up. Visually 
verify that aerosol is coming out of the 
mouthpiece or mask to indicate that the 
device is functioning properly.
NOTE: (1) If you are using batteries and the green light is flash-
ing, your batteries are low. You may finish your treatment, and 
7
12
DAILY: (Cont.)
1. Wash the Nebulizer Unit (Medication Cup and Cap, Nebulizer 
Body, and Mouthpiece or Mask Adapter and Mask) by hand 
in a solution of warm water and mild liquid dishwashing 
soap. Use soap intended only for washing dishes by hand.
Do not use dishwasher detergent as it is too abrasive. Do not 
leave parts soaking in soapy water for more than 20 minutes.
2. Rinse all pieces thoroughly under running hot tap water.
Make sure that no soap is left on any of the parts.
3. Shake off excess water and allow parts to air dry on a clean,
dry towel, out of the reach of children.
4. Make certain that all cleaned parts are completely dry before 
you store or use them. Do NOT remove the retaining ring 
under the medication cup.
NOTE: If the Medication Cup is not properly cleaned after every
use the nebulizer may become clogged with medication.
Improper cleaning will result in poor performance and increased
time to nebulize the medication.
NOTE: Water remaining on the underside of the Aeroneb
®
Go
generator after cleaning may cause the unit not to work. Be sure
the unit is completely dry prior to use.
NOTE: Do NOT clean the nebulizer in a dishwasher.
OPTIONAL DISINFECTION:
If you would like to disinfect your Aeroneb
®
Go Nebulizer, you
may use the following method.
VINEGAR SOLUTION:
I. First, clean the Nebulizer Unit as described in the daily 
cleaning section, following steps 1- 2.
II. Fill a container with a mixture of three (3) cups hot water to 
one (1) cup white distilled vinegar.
(example: Heinz Distilled White Vinegar)