User's Guide

23
Using the Monitor
Using the Monitor
Once connected to the pad, and with no patient/resident on the pad, the
monitor is ready for a patient/resident to get on the pad.
When a patient/resident has been detected on the pad, the Reset Button
flashes green for 3 seconds before arming, allowing time for the
patient/resident to shift around slightly until a comfortable position is
achieved. The arming is indicated by a Positive Confirmation Tone.
Any exit from the pad during the first 3 second flash period cancels the
arming process.
After arming, the monitor sounds a Pad alarm on a pad exit.
Note: The Floor Pad has no arming stage and alarms immediately upon
weight being applied.
Hold Mode (30 Seconds)
Hold Mode is used to allow the caregiver 30 seconds to remove a
patient/resident from a monitored pad without triggering an alarm. If the
patient/residents weight is applied to the pad (3 seconds of constant weight
pressure on the pad) while in Hold mode, normal monitoring resumes. A positive
tone is heard and the amber blinking stops.
1. While the patient/resident is being monitored on the pad, press and
hold the Reset Button on the monitor for 2 seconds*. An audible
indication “Hold mode 30 seconds” is announced.
2. A 30 second timer is set. The patient/resident can now be
removed from the pad without triggering an alarm.
3. The Reset Button flashes amber once per second for 30
seconds. When Hold mode is over, an audible positive
indication is announced.
4. To move out of Hold mode before the 30 seconds has passed, press
and hold the Reset Button for 2 seconds, or return the
patient/resident to the pad. The Reset Button stops blinking, followed
by a Positive Confirmation Tone. If there is weight applied to the pad -
the monitor will arm.
5. When the timer runs out, the monitor resumes normal
operation.