35
Guidance and manufacturer’s Declaration–electromagnetic
emission
SW-1000Sscooteris intended for use in the electromagnetic
environment specified below. The customer or the user of the
SW-1000Sscooter should assure that it is used in such an
environment.
Immunity test
IEC 60601
test level
Complia
nce
level
electromagnetic
environment-guidanc
e
Electrostatic
discharge
ESD
IEC 61000S-4-2
±6 kV contact
±8 kV air
±6 kV
contact
±8 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical fast
transient/ burst
IEC 61000S-4-4
±2kV for power
supply lines
±1kV for input/
output lines
±2kV for
power
supply
lines
±1kV for
input/
output
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000S-4-5
±1 kV
differential
mode
±2 kV common
mode
±1 kV
differenti
al mode
±2 kV
common
mode
Mains power quality
should be that of a
typical commercial or
hospital environment.