User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
118
2900H EN 20071108Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
Print head profile calibration
To start the manual print head profile calibration:
1 Perform step 1 to 2 of the ‘Accessing the calibration menu’ procedure on page 111
to select the ‘Calibration’ screen.
2 Press the Down key once to select ‘Printhead profile’, and press the Confirm key.
The ‘Select input tray’ screen appears:
3 Press the Up/Down arrow keys to select an input tray, followed by the Confirm
key (e.g. choose upper input tray).
The system will start with Phase 1 of the calibration: Registration:
As a result, the ‘Registration’ print out appears.
Print head profile calibration must be done whenever the profile of the
print head needs to be improved (manual print head profile calibration).
For more information on maintaining image quality, refer to ‘Maintaining
image quality and resolving Image quality problems’ on page 220.
In the procedure, we will assume a print head profile calibration on the upper input
tray. The procedure for the lower input tray is identical.