User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
119
2900H EN 20071108 Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
4 After registration, the system will continue with Phase 2 of the calibration: the
print head profile calibration:
As a result, the ‘TPH profile’ print out appears.
5 In case the print head profile calibration has failed, the following message
appears:
Press the Confirm key to retry the calibration, or the Escape key to quit.
6 In case the film calibration has completed successfully, the following message
appears:
• The percentage for the ‘Registration compensation’ reflects the degree to which the
system had to compensate for incorrect horizontal positioning of the built-in
densitometer (100% compensation = maximal correction).
• The percentage for the ‘print head profile compensation’ reflects the degree to
which the system had to compensate for density variations in the print head profile
(100% compensation = maximal correction).
Both values are important for the service technician.
When the calibration process is interrupted, you will have to start the
calibration procedure again. Otherwise, the old calibration parameters
will be used.
Blinking
Blinking