User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
92
2900H EN 20071108Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
5 Press the Up/Down arrow keys to select the preferred film calibration default.
When the setting is ON, the Drystar 5500 will ask you to enter the calibration
frequency:
Enter the calibration frequency by means of the Up/Down arrow keys. Refer to ‘Data
entry’ on page 33.
6 Press the Confirm key to store the data, or the Escape key to quit the procedure
without any changes.
1 pack = 100 sheets
E.g., if the frequency is set to 5, this means that a film calibration is executed
automatically every 5 packs or every 500 sheets.
You can specify to perform an automatic film calibration after loading every “n“
film packs.
For more information on film calibration settings: refer to ‘Understanding the
calibration policy and the calibration initiation’ on page 221.
5