User Manual

Table Of Contents
92
2900H EN 20071108Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
5 Press the Up/Down arrow keys to select the preferred film calibration default.
When the setting is ON, the Drystar 5500 will ask you to enter the calibration
frequency:
Enter the calibration frequency by means of the Up/Down arrow keys. Refer to ‘Data
entry’ on page 33.
6 Press the Confirm key to store the data, or the Escape key to quit the procedure
without any changes.
1 pack = 100 sheets
E.g., if the frequency is set to 5, this means that a film calibration is executed
automatically every 5 packs or every 500 sheets.
You can specify to perform an automatic film calibration after loading every “n“
film packs.
For more information on film calibration settings: refer to ‘Understanding the
calibration policy and the calibration initiation’ on page 221.
5