User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
97
2900H EN 20071108 Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
6 On the ‘Image quality’ screen press the Down key once, followed by the Confirm
key to select ‘Interpolation’.
The ‘Interpolation’ screen for the selected modality is displayed:
The current interpolation setting for the selected modality is marked with an *.
7 Press the Up/Down arrow keys to select the desired Interpolation setting for
this modality.
• If you select “Cubic Hi-Res” the following screen will appear, allowing you to adjust
the smoothing factor:
• If you select “CubicBell” the following screen will appear, allowing you to adjust
the smoothing factor:
8 Press the Confirm key to store the data, or the Escape key to quit the procedure
without any changes.
The user should be aware that the interpolation values differ when the
A#Sharp feature is deactivated. The A#Sharp feature, which is a
technology that enhances the image sharpness, is active by default. For
more information about (de)activating the A#Sharp feature, refer to the
Drystar 5500/5503 Service manual.
<Modality name>
1 CubicBell
2*Cubic Hi-Res
3 Linear
4 Replication
5 None
<Modality name>
Default 0.70
<Modality name>
Default 0.90