User's Manual Part 2
Table Of Contents
- Table of contents
- Introducing the Drystar5300
- Basic operation (Operator mode)
- Advanced operation (Key-operator mode)
- Overview of key-operator functions
- Viewing printer information
- Accessing the Show settings menu
- Viewing general information
- Viewing input tray information
- Viewing network (DICOM) information
- Viewing general image quality information
- Viewing default densitometer information
- Viewing image quality information of a modality
- Viewing installed data-sets information
- Viewing logical printers information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Maintaining optimal image quality
- Service actions
- Quality control procedure
- Installation operations
- Controlling the Drystar5300 via a remote PC (with browser)
- System description
- Preventive maintenance
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Index
105
2920B EN 20031206
4 When the film calibration has been completed successfully, three cases are
possible:
• Calibration successful, results OK.
The obtained MaxD (3.10) after calibration, is displayed:
(xxxxx) is the code of the Densitometer type (refer to ‘Changing the default densitometer’
on page 86).
• Calibration successful, but MaxD differs more than 2% from the target value.
The obtained MaxD (3.10) after calibration, is displayed:
(xxxxx) is the code of the Densitometer type (refer to ‘Changing the default densitometer’
on page 86).
• Calibration successful, but the maximum density has been reached.
The obtained MaxD (3.10) after calibration, is displayed:
(xxxxx) is the code of the Densitometer type (refer to ‘Changing the default densitometer’
on page 86).
5 Press the Confirm key to return to the Calibration menu.
When the calibration process is interrupted, you will have to start the
calibration procedure again. Otherwise the old calibration parameters
will be used.
Film
calibration OK
MaxD = 3.10 (xxxxx)
CA
Flashing
Flashing
Retry
calibration or
Resume printing
MaxD = 3.10 (xxxxx)
CA
Flashing
Flashing
Film
calibration
MaxD is reached
MaxD = 3.10 (xxxxx)
CA
Flashing
Flashing