" The calculated ‘Operating levels’ should respect the following tolerances:
" The tolerances may be different according to local regulations.
10 These charts will be used for the daily quality test. For more information, refer
to ‘Performing the daily QC test’ on page 54.
Operating level Tolerance
Base + Fog - 0.03 / + 0.05
Mid Density +/- 0.20
DD (Density Difference) +/- 0.20
MaxD +/- 0.25