Technical data

1100 Series DAD and MWD Reference Manual 123
Repairing the Detector 4
3 Unscrew the capillaries from the flow cell. DO NOT use the
adapter at this time!
4 Using for example a toothpick, press on the plastic part
and slide the quartz body out of the cell housing.
5 The quartz body and the cell seal assembly can be
separated for cleaning purpose.
6 This figure shows the correct holding of the quartz body
and the cell seal assembly.