Technical data

1100 Series DAD and MWD Reference Manual 129
Repairing the Detector 4
5
Replace the flow cell cover and fix the six screws. Next steps:
Perform a holmium oxide test, see “Holmium Oxide
Te s t on page 86 to check the proper function of the
holmium oxide filter.
Insert the flow cell, see “Exchanging a Flow Cell” on
page 105.
Replace the front cover.
Turn on the flow.