Technical data
Chapter 5 103
Spectrum Analyzer
Using the Analyzer as a Fixed Tune Receiver
Spectrum Analyzer
and turn AM on.
Step 3. Select the spectrum analyzer mode:
Press 
Mode, Spectrum Analyzer.
Step 4. Preset the analyzer.
Press 
Mode Preset.
Step 5. Set the center frequency, span, RBW and the sweep time.
Press 
FREQ Channel, Center Frequency, 300, MHz.
Press 
SPAN X Scale, Span, 500, kHz.
Press 
BW, Res BW, 30, kHz.
Step 6. Set the y-axis units to volts:
Press 
AMPTD Y Scale, More, Y-Axis Units, Volts. 
Step 7. Position the signal peak near the reference level:
Press 
AMPTD Y Scale, Ref Level, (rotate front-panel knob).
Step 8. Change the y-scale type to linear:
Press 
AMPTD Y Scale, Scale Type (Lin).
Step 9. Set the analyzer in zero span to make time-domain measurements:
Press 
SPAN X Scale, Zero Span.
Press 
Control/Sweep, Sweep Time, 5, ms.
Step 10. Use the video trigger to stabilize the trace:
Press 
Meas Setup, Trigger, Video. Adjust the trigger level for a stable 
trace.
Since the modulation is a steady tone, you can use video trigger to 
trigger the analyzer sweep on the waveform and stabilize the trace, 
much like an oscilloscope. See Figure 5-26. 
NOTE If the trigger level is set too high or too low when video trigger mode is 
activated, the sweep stops. You need to adjust the trigger level up or 
down with the front-panel knob until the sweep begins again.
Step 11. Measure the AM rate using delta markers: 
Press 
Peak Search, Marker, Delta, Peak Search, Next Pk.
Use markers and delta markers to measure the AM rate. Place the 
marker on a peak and then use a delta marker to measure the time 