Technical data
118 Chapter 7 
Stimulus Response Measurements (Option N8995A)
Two Port Insertion Loss
Stimulus Response Measurements 
(Option N8995A)
Step 7. Connect the cable (but not the DUT) from the tracking generator output 
to the analyzer input as shown in Figure 7-1. 
Figure 7-1 Two Port Insertion Loss Normalization Test Setup 
Step 8. Normalize the frequency response:
Press 
FREQ Channel, Normalize and follow the instructions on the Normalize 
Wizard.
Step 9.
To measure the rejection of a low pass filter:
Connect the DUT between the RF Input and RF Output of the analyzer 
as shown in Figure 7-2. 
Note that the units of the reference level are dB, indicating that this is 
a relative measurement.
This example uses a 50 MHz low pass filter as the DUT. 
Figure 7-2 Two Port Insertion Loss Measurement Test Setup 
Step 10. Place the reference marker at the specified cutoff frequency:
Press 
Marker, Normal, 50, MHz.
Step 11. Place the second marker at 100 MHz:
Press 
Delta, 50, MHz.










