Technical data

Chapter 7 119
Stimulus Response Measurements (Option N8995A)
Two Port Insertion Loss
Stimulus Response Measurements
(Option N8995A)
In this example, the attenuation over this frequency range is
60.7 dB/octave (one octave above the cutoff frequency).
Step 12. Use the front-panel knob to place the marker at the highest peak in the
stop band to determine the minimum stop band attenuation. In this
example, the peak occurs at 90.47 MHz. The attenuation is 57.6 dB.
Figure 7-3 Minimum Stop Band Attenuation