Technical data
120 Chapter 7 
Stimulus Response Measurements (Option N8995A)
One Port Insertion Loss
Stimulus Response Measurements 
(Option N8995A)
One Port Insertion Loss
The one port insertion loss measurement allows you to quantify signal 
loss in a cable or other device without connecting both ends to the 
analyzer. This measurement can be especially useful in measuring the 
loss of a feedline connected to the antenna on a tower. 
This measurement is less accurate than two port insertion loss, 
however. Therefore, when it’s practical to connect both ends of a device 
to the analyzer—for example, for short cables or attenuators—it is 
better to use two port insertion loss.
NOTE Test signals can cause interference. When testing cables attached to 
antennas, test signals are radiated. Verify that the signal used for the 
test cannot cause interference to another antenna. 
NOTE The One Port Insertion Loss calibration is the same calibration as 
performed for two other measurements: Return Loss and Distance to 
Fault (as long as you use the manual frequency method). If you make 
the calibration for any of these three measurements, the calibration will 
apply to the other two measurements and “Calibrated” will be displayed 
on the screen for all three. 
The calibration remains valid until you power off the analyzer or 
change the start or stop frequency. 
Step 1. Connect the calibrating devices to the analyzer RF Input when 
prompted in the procedure, as shown in Figure 7-4, or in the calibration 
wizard. 
NOTE DO NOT make the connection at this time. You will be directed when to 
make the connections later in the procedure.
This example uses a 15 foot cable as the DUT. 










