Technical data
Chapter 7 121
Stimulus Response Measurements (Option N8995A)
One Port Insertion Loss
Stimulus Response Measurements 
(Option N8995A)
Figure 7-4 One Port Insertion Loss Measurement
Step 2. Set the analyzer to the One Port Insertion Loss measurement:
Press 
Mode, Stimulus/Response, One Port Insertion Loss
Step 3.
Preset the analyzer:
Press 
Mode Preset, Meas, One Port Insertion Loss.
Step 4. Set the start and stop frequencies:
Press 
FREQ Channel, Start Freq, 100, MHz.
Press 
FREQ Channel, Stop Freq, 2, GHz.
Step 5. Turn averaging off:
Press 
Meas Setup, Avg Mode, Off.
Step 6. Calibrate the measurement:
Press 
FREQ Channel, Calibrate and follow the instructions on the Calibration 
Wizard. The analyzer will calibrate over the desired frequency range.
Step 7.
Connect the DUT to the analyzer, as described in step 1. Note that the 
units of the reference level are dB, indicating that this is relative 
measurement.
Step 8. Change the amplitude scale to 2 dB per division:
PRess 
AMPTD Y Scale, Scale/Div, 2, dB.
Step 9. Place a marker on the results at the frequency of interest. In this 
example, the marker is placed at 986.667 MHz. As you can see the loss 
is 2.1 dB.










