Technical data
Chapter 7 123
Stimulus Response Measurements (Option N8995A)
Return Loss
Stimulus Response Measurements 
(Option N8995A)
Return Loss
Return loss is a measure of reflection characteristics. One way you can 
use the return loss measurement is to detect problems in an antenna 
feedline system or the antenna itself. A portion of the incident power 
will be reflected back to the source from each transmission line fault as 
well as the antenna. The ratio of the reflected voltages to the incident 
voltage is called the reflection coefficient. The reflection coefficient is a 
complex number, meaning it has both magnitude and phase 
information. In S-parameter terms, Return Loss is referred to as an S
11
measurement.
NOTE Test signals can cause interference. When testing cables attached to 
antennas, test signals are radiated. Verify that the signal used for the 
test cannot cause interference to another antenna. 
NOTE The Return Loss calibration is the same calibration as performed for 
One Port Insertion Loss and Distance to Fault (as long as you use the 
manual frequency method). If you make the calibration for any of these 
three measurements, the calibration will apply to the other two 
measurements and “Calibrated” will be displayed on the screen for all 
three. 
The calibration remains valid until you power off the analyzer or 
change the start or stop frequency. 
Step 1. Set the analyzer to the Stimulus/ Response Mode and the Return Loss 
measurement:
Press 
Mode, Stimulus/Response, Return Loss
Step 2.
Preset the analyzer:
Press 
Mode Preset, Meas, Return Loss.
Step 3. Set the start and stop frequencies:
Press 
FREQ Channel, Start Freq, 10, MHz.
Press 
FREQ Channel, Stop Freq, 250, MHz.
Step 4. Turn averaging off:
Press 
Meas Setup, Averaging, Off.
Step 5. Calibrate the measurement:
Press 
FREQ Channel, Calibrate and follow the instructions on the 
Calibration Wizard. The analyzer will calibrate over the desired 










