Technical data
Chapter 7 125
Stimulus Response Measurements (Option N8995A)
Distance to Fault
Stimulus Response Measurements 
(Option N8995A)
Distance to Fault
A signal is transmitted from the RF Output connector of the analyzer to 
the cable-under-test. The signals reflected from faults in the cable are 
received by the analyzer.
In performing this measurement, the analyzer uses frequency domain 
reflectometry. The changing interference of the transmitted and 
reflected signals contains information about the distance to one or more 
faults. This information can be used to find the physical distance to the 
faults. The distance displayed on the analyzer is the physical distance 
to the probable faults, corrected for the cable loss and velocity factor of 
the cable.
The analyzer provides two ways of measuring distance to fault:
• Manual Frequency Range. You select the start and stop 
frequencies, which define the measured distance. Generally, the 
typical start and stop frequencies you use will result in a measured 
distance that will be larger than the distance over which you want to 
look for faults. To help isolate faults over the length of interest, you 
can set a displayed distance less than the measured distance. The 
displayed distance is set using the 
Start Distance and Stop Distance 
menu keys on the [
Freq/Dist/Calibrate] menu. Keep in mind that there 
are 256 measurement points across the measured distance. 
Therefore, the measurement points across the chosen displayed 
length will be a ratio of displayed distance to measured distance 
times 256. The higher the ratio, the less measurement resolution. In 
most cases, the resolution will be adequate to determine the faults, 
but if more resolution is needed you can increase the span between 
the start and stop frequencies (which will decrease the measured 
distance) or use the other approach, automatic frequency range. If 
the measurement distance is not long enough for the cable you are 
testing, reduce the span between the start and stop frequencies 
(which will increase the measurement distance) or use automatic 
frequency range. 
• Automatic Frequency Range. You select the measurement 
distance, and the analyzer automatically selects the start and stop 
frequencies. This measurement distance is set using the 
Start 
Distance
 and Stop Distance menu keys on the [Freq/Dist/Calibrate] 
menu. In this mode, the displayed and measured differences are the 
same. There are 256 measurement points across the distance you 
set. This approach provides the maximum measurement resolution 
across the selected distance. The disadvantage is that the start and 
stop frequencies are automatically set and may limit the analyzer's 
ability to sweep through filters or lighting protectors. This mode is 
best used for checking a cable that has no frequency limiting devices. 










