Stimulus Response Measurement Concepts
Concepts
level are dB, indicating that this is a relative measurement.
To make a reflection measurement accurately, it is necessary to perform 
an open/short/load calibration. An open, short, and load are included in 
the Stimulus Response Calibration Kit, Option SRK.
Press 
Mode, Stimulus/Response, Return Loss. Set the desired start and 
stop frequencies. Press 
Calibrate and follow the instructions.
After the calibration is complete, connect the DUT to the RF OUTPUT 
connector to make your return loss measurement. The marker readout 
returns the amplitude values in both return loss and VSWR.