Technical data
152 Chapter 10 
Basic System Operations
Setting System References
Basic System Operations
Setting System References
The Agilent CSA provide a utility to preconfigure the global settings for 
the box.
Selecting a Frequency/Timing Reference
Perform this procedure to select a common frequency or timing 
reference to be used for all measurement tools (when applicable). 
1. Press System, Freq/Time Reference
2.
Using the knob or the up/down arrow navigation keys to highlight 
the frequency/timing reference you want.
3. Press Select.
NOTE A frequency/time reference indicator in the lower-right of the screen 
shows both the selected reference and its status. 
Reference indicators include: Int Ref, Even Sec, Ext 1.0 MHz, 
Ext 2.048 MHz, Ext 4.95 MHz, Ext 10 MHz, Ext 13 MHz, Ext 15 MHz, 
or Ext 19.66 MHz. 
Status indicators include: 
• Green dot to indicate that the reference is locked
• Yellow triangle to indicate that the reference is acquiring lock
• Red X to indicate that the reference is not locked