CAUTION The external power supply has autoranging line voltage input. Be sure 
the supply voltage is within the specified range. (Refer to the 
specifications guide for your analyzer.)
CAUTION When operating this product with the external power supply, always 
use the three-prong power cord supplied with this product. Failure to 
ensure adequate earth grounding by not using this cord can cause 
product damage.
CAUTION VENTILATION REQUIREMENTS: When installing the product in a 
cabinet, the convection into and out of the product must not be 
restricted. The ambient temperature (outside the cabinet) must be less 
than the maximum operating temperature of the product by 4°C for 
every 100 watts dissipated in the cabinet. If the total power dissipated 
in the cabinet is greater than 800 watts, then forced convection must be 
used. 
Lifting and Handling
When lifting and handling the Agilent N1996A Spectrum Analyzer use 
ergonomically correct procedures. If so equipped, lift and carry the 
analyzer by the bail handle. 