Technical data
Chapter 5 83
Spectrum Analyzer
Measuring Multiple Signals
Spectrum Analyzer
Figure 5-8 Unresolved Signals of Equal Amplitude 
Step 4. Change the resolution bandwidth (RBW) to 100 kHz so that the RBW 
setting is less than or equal to the frequency separation of the two 
signals:
Press 
BW, Res BW (Manual), 100, kHz.
Notice that the peak of the signal has become flattened indicating that 
two signals may be present. 
Figure 5-9 Resolving Signals of Equal Amplitude 
As the resolution bandwidth is decreased, resolution of the individual 
signals is improved and the sweep time is increased. For fastest 
measurement times, use the widest possible resolution bandwidth. 
Under factory preset conditions, the resolution bandwidth is “coupled” 
(or linked) to the span. 
Since the resolution bandwidth has been changed from the coupled 
value, a # mark appears next to Res BW in the lower-left corner of the 
screen, indicating that the resolution bandwidth is uncoupled. (For 
more information on resolution bandwidth, refer to the 
Res BW 