Specifications
Chapter 12 105
Using Segmented Sweep (ESA-E Series Spectrum Analyzers)
Measuring Harmonics Using Segmented Sweep
Using Segmented Sweep (ESA-E 
Series Spectrum Analyzers)
Press Center Freq, 150, MHz.
Press 
Span, 100, kHz.
Press 
Video BW, 30, Hz.
Notice that the total sweep time for the 3 segments is only 8.13 seconds 
(plus a minimal time for transitions between segments). This is 
considerably shorter than the 193.3 seconds required to view the same 
signals without segmented sweep.
Figure 12-2 Reducing Sweep Time with Segmented Sweep