Specifications
106 Chapter 12 
Using Segmented Sweep (ESA-E Series Spectrum Analyzers)
Using Segmented Sweep With Limit Lines
Using Segmented Sweep (ESA-E 
Series Spectrum Analyzers)
Using Segmented Sweep With Limit Lines
Segmented sweep can also use other standard spectrum analyzer 
functionality, such as limit lines. In this procedure three segments are 
used. In the first segment, the limit lines are used to verify that the 
signal phase noise is acceptable. The second and third segments are 
used to verify the harmonic signals are below a specified amplitude.
Step 1. Connect the 10 MHz REF OUT from the rear panel to the front-panel 
INPUT. Preset the instrument:
Press 
Preset, Factory Preset (if present). 
Step 2. Place the peak of the fundamental signal at the top of the graticule:
Press 
AMPLITUDE Y Scale, Ref Level, 8, dBm.
Step 3. Open the segmented sweep editor:
Press 
Sweep, Segmented, Modify, Edit.
Step 4. Display the fundamental signal in the first segment:
Press 
Segment, 1, Enter.
Press 
Center Freq, 10, MHz.
Press 
Span, 500, kHz.
Press 
Res BW, 3, kHz.
Press 
Video BW, 30, Hz.
Press 
Points, 1000, Enter.
Step 5. Display the second harmonic in the second segment:
Press
 Segment, 2, Enter.
Press 
Center Freq, 20, MHz. 
Press 
Span, 100, kHz.
Step 6. Display the third harmonic in the third segment:
Press 
Segment, 3, Enter.
Press 
Center Freq, 30, MHz. 
Press 
Span, 100, kHz.
Step 7. Open the limit line editor:
Press 
Display, Limits, Limit 1, Type (Upper), Edit.
Step 8. Enter the limit line data from Table 12-1 below. To enter the table data:
Press 
Point, 1, Enter, Frequency, 9.75, MHz, Amplitude, −80, dBm, 
Connected (Yes).
Step 9. Begin the next limit line point:
Press 
Tab ⇒|. 