Specifications
Chapter 2 13
Measuring Multiple Signals
Comparing Signals on the Same Screen Using Marker Delta
Measuring Multiple Signals
Step 4. Place a marker at the highest peak on the display (10 MHz): 
Press 
Peak Search.
The 
Next Pk Right and Next Pk Left softkeys are available to move the 
marker from peak to peak. The marker should be on the 10 MHz 
reference signal:
Step 5. Anchor the first marker and activate a second marker:
Press 
Marker, Delta.
The label on the first marker now reads 1R, indicating that it is the 
reference point.
Step 6. Move the second marker to another signal peak using the front-panel 
knob or by using the 
Peak Search key: 
Press 
Peak Search, Next Peak or
Press 
Peak Search, Next Pk Right or Next Pk Left.
The amplitude and frequency difference between the markers is 
displayed in the active function block. For ESA see the left side of 
Figure 2-2 and the right side for PSA.
Figure 2-2 Using the Delta Marker Function (ESA left, PSA right)
NOTE The resolution of the marker readings can be increased by turning on 
the frequency count function.