Specifications
Chapter 15 143
Concepts
Time Gating Concepts
Concepts
View/Trace (ESA) or Trace/View (PSA) menu) to determine the frequency 
of peak energy.
To optimize measurement speed, set the span narrow enough so that 
the display will still show the signal characteristics you want to 
measure. For example, if you wanted to look for spurious signals within 
a 200 kHz frequency range, you might set the frequency span to just 
over 200 kHz.
Step 4. Determine the setup time and signal delay to set up the gate signal. 
Turn on the gate and adjust the gate parameters including gate delay 
and gate length as shown below. 
Generally, the gate should be positioned over a part of the signal that is 
stable, not over a pulse edge or other transition that might disturb the 
spectrum. Starting the gate at the center of the pulse gives a setup time 
of about half the pulse width. Setup time describes the length of time 
during which that signal is present and stable before the gate comes on. 
The setup time (SUT) must be adequately long enough for the RBW 
filters to settle following the burst-on transients. Signal delay (SD) is 
the length of time after the trigger, but before the signal of interest 
occurs and becomes stable. If the trigger occurs simultaneously with the 
signal of interest, SD is equal to zero, and SUT is equal to the gate 
delay. Otherwise, SUT is equal to the gate delay minus SD. See Figure 
15-16.
Figure 15-16  Positioning the Gate 
There is flexibility in positioning the gate, but some positions offer a 
wider choice of resolution bandwidths. A good rule of thumb is to 
position the gate from 20% to 80% of the burst for PSA, and 25% to 80% 
of the burst for ESA. Doing so provides a reasonable compromise 
between setup time and gate length.










